Ratfisch@compuserve.com


Startpage RS 53-T RS 55-T RS 555 RS 2400 RS 11 Peripherals
Homepage Tar Analzer TA120 Representatives


++New Product++++New Product++++New Product++++New Product++++New Product++++New Product++

Quasi-continuous Online Measurements of Tar Content

of Raw Gases from Biomass Gasification

with the

Tar Analyzer

TA 120-3


++New Product++++++New Product++++++New Product++++++New Product++++++New Product++

Sponsored by

[Front View TA 120-3 (25 kB)]


General:



Basic Functions:





Specifications:


Detector Flame Ionization Detector (FID)
Analytical Temperature 300 °C
Filter Temperatures 0 to +100 °C (1°C-increments)
Number of Filters 2
Numbers of Sample Inputs 2
Sample Line Temp.-Control 2, (Preset to 300°C)
Measurement Ranges (MR) 3 (12, 60, 120gC/m3)
Measurement Cycle 120 seconds
Limit of Detection < 0,2 % of THC
Reproducibility +/- 1 % of Value
Linearity +/- 2 % of MR
Sample Pumps 3, heated Venturi
Sample Gas Flow 100 - 300 l/h
Warm-up approx. 60 Minutes
Temperature Control PID-Controller
Ambient Temperature +20 to +35 °C
PC-Interface RS 485/232 Converter
Alarm Functions
  • Flame Monitoring
  • Sample Valve
  • Analytical Temperature
  • Filter Temperature (2)
  • Sensor Conditions
Status Information
  • Analytical Status
  • Sample Input
  • Analytical Temperature
  • Filter Temperature (2)
Software TA 120 DataAquisition© for PC/Laptop under WINDOWS
Operational Gases
  • Fuel FID:
    Hydrogen (Purity 5.0), 2 bar, approx. 2 l/h
  • Combustion Air FID:
    Compressed Air (HC-free), 3 bar, approx. 20 l/h
  • Carrier Gas:
    Nitrogen (Purity 5.0), 2 bar, approx. 5 l/h
  • Compressed Air:
    For Valve Control, 6 bar, approx. 1000 l/h
  • Calibration Gas:
    approx. 5 Vol.-% Methane/Propane in Nitrogen, approx. 2 bar
Gas Connections 1/4" Clamping Ring Screw Nut
Power 230 V / 50 Hz / 900 VA
Dimensions 6 HE x 19" x 480 mm (H x W x D)
Weight approx. 30 kg